Authors: Remy Lobo ( University of Michigan Health Systems) , Ashley Harrie ( University of Michigan Medical School) , Diana M Gomez-Hassan ( University of Michigan Health Systems) , Maria Masotti ( University of Michigan Health Systems) , David Seidenwurm ( Sutter Health Sacramento) , Ashok Srinivasan ( University of Michigan Health Systems)

  • Quantifying Neuroquant: Comparison of Outputs with Software Updates

    Abstract Articles

    Quantifying Neuroquant: Comparison of Outputs with Software Updates

    Authors: , , , , ,

Download

Downloads:
Download HTML

41099b72-8e43-4744-a1e5-d9f9e249858d

Introduction

NeuroQuant (NQ) is an automated brain segmentation software that aims to improve diagnostic accuracy and reduce variability between diagnostic radiologists for various neurodegenerative diseases. There have been multiple NQ software updates in the past five years; this project evaluates consecutive patient series between two recent updates (v.2 and v4 as well as v3 and v4) to determine the variability in quantitative outputs and normative data.

Methods

11 consecutive patients previously scanned with our institution’s routine dementia protocol had images processed with NQ v2.3 and repeat analysis performed with the NQ v4 software package. The same process was executed with 39 scans previously processed with NQ v3.0.1. Outputs from 34 different variables (e.g. hippocampal percentile, cortical gray volume) were compared between the two software packages using paired t-tests of median values assessing for variance amongst interquartile ranges. Additionally, Bland-Altman plots were generated for total volumes of each lobe, percentile, cortical gray volume, hippocampal volume and percent to look for variance between the datasets.

Results

The results show overall high correlation between hippocampal volumes/percent (p-value 0.001 to 0.016), in both paired output comparisons. Cortical gray volume percentiles had a higher p-value of 0.1 on paired t-test. The Bland-Altman plot of hippocampal percentile values had a standard deviation of bias at 9 for the most recent software comparison (v.3.0.1 and v.4). The cortical gray percentile output had a standard deviation of bias of 17. Whole brain raw volumes and percentiles also showed variance between both cohorts, with a standard deviation up to 13.

Conclusions

Overall, interquartile variance was highly correlated in both comparisons. Though data correlates well between the software packages, standard deviations of bias approach 9, which can place patients' values beyond two standard deviations of expected controls, impacting their clinical diagnosis. Assessment of whole brain volume, including cortical gray volume showed higher variation in bias; these outputs should be used with greater caution. While automated software packages such as NeuroQuant have useful application in the clinical evaluation of dementia, this project draws attention to the variance between data outputs and how such variations have the potential to impact clinical decision making.

References

  1. Lee JY, Oh SW, Chung MS, Park JE, Moon Y, Jeon HJ, Moon WJ. Clinically Available Software for Automatic Brain Volumetry: Comparisons of Volume Measurements and Validation of Intermethod Reliability. Korean J Radiol. 2021 Mar;22(3):405-414. doi: 10.3348/kjr.2020.0518. Epub 2020 Nov 3. PMID: 33236539; PMCID: PMC7909859.

  2. Ochs AL, Ross DE, Zannoni MD, et al. Alzheimer's Disease Neuroimaging Initiative. Comparison of Automated Brain Volume Measures obtained with NeuroQuant and FreeSurfer. J Neuroimaging. 2015 Sep-Oct;25(5):721-7. doi: 10.1111/jon.12229. Epub 2015 Feb 26. PMID: 25727700.

  3. Persson K, Leonardsen EH, Edwin TH, et al. Diagnostic accuracy of brain age prediction in a memory clinic population and comparison with clinically available volumetric measures. Sci Rep. 2023 Sep 11;13(1):14957. doi: 10.1038/s41598-023-42354-0. PMID: 37696909; PMCID: PMC10495330.

  4. Persson K, Barca ML, Cavallin L, Brækhus A, Knapskog AB, Selbæk G, Engedal K. Comparison of automated volumetry of the hippocampus using NeuroQuant® and visual assessment of the medial temporal lobe in Alzheimer's disease. Acta Radiol. 2018 Aug;59(8):997-1001. doi: 10.1177/0284185117743778. Epub 2017 Nov 27. PMID: 29172642.

  5. Persson K, Selbæk G, Brækhus A, Beyer M, Barca M, Engedal K. Fully automated structural MRI of the brain in clinical dementia workup. Acta Radiol. 2017 Jun;58(6):740-747. doi: 10.1177/0284185116669874. Epub 2016 Sep 28. PMID: 27687251.

  6. Tanabe J, Lim MF, Dash S, Pattee J, Steach B, Pressman P, Bettcher BM, Honce JM, Potigailo VA, Colantoni W, Zander D, Thaker AA. Automated Volumetric Software in Dementia: Help or Hindrance to the Neuroradiologist? AJNR Am J Neuroradiol. 2024 Nov 7;45(11):1737-1744. doi: 10.3174/ajnr.A8406. PMID: 39362700; PMCID: PMC11543079.

  7. Tanpitukpongse TP, Mazurowski MA, Ikhena J, Petrella JR; Alzheimer's Disease Neuroimaging Initiative. Predictive Utility of Marketed Volumetric Software Tools in Subjects at Risk for Alzheimer Disease: Do Regions Outside the Hippocampus Matter? AJNR Am J Neuroradiol. 2017 Mar;38(3):546-552. doi: 10.3174/ajnr.A5061. Epub 2017 Jan 5. PMID: 28057634; PMCID: PMC5352470.